Figure 4-3
13 Press the Prev key.
14 Press the Fill Table From SNS menu key. Wait until the data is uploaded from
15 Verify the data has been transferred from the SNS to the NFA. The data
Keysight N4000/1/2A Operating and Service Manual
A typical cleared ENR table
the SNS before proceeding.
transfer should be similar to Figure 4-4 on page 34. If it is not, return the SNS
to Keysight Technologies for repair.
Operating Guide
4
33