Keysight Technologies N4000A SNS Series Operating And Service Manual page 38

Smart noise sources
Table of Contents

Advertisement

4
Operating Guide
3 Connect the SNS to the NFA Port using the multi-pin 11730 cable, as shown in
Figure 4-9
4 Press the Frequency/Points key.
5 Press the Frequency Mode menu key.
6 Press the Fixed menu key.
7 Press the Fixed Frequency menu key. Enter 1.0 GHz as a fixed frequency value,
8 Press the Sweep key.
9 Press the Manual Meas menu key.
10 Press the Manual State menu key and set it to Manual State (On).
11 Press the Noise Source menu key and set it to Noise Source (On).
12 Press the IF Att menu key and set it to IF Att (Hold).
13 Monitor the power value which appears in the manual measurement screen.
38
Figure
4-9.
SNS connected to NFA graphic
by pressing the 1 numeric key followed by pressing the GHz menu key.
Figure 4-10 on page
39
shows a typical example of this.
Keysight N4000/1/2A Operating and Service Manual