ProMinent DULCOMETER D1C Operating Instructions Manual page 22

Measured variable chlorine dioxide
Table of Contents

Advertisement

Complete Operating Menu / Description
Standard Signal Output 1
Access to all setting menus can be blocked with an access code !
Only with standard signal output
mA output 1
setting ?
Standard Signal Output 2
Access to all setting menus can be blocked with an access code !
Only with 2 standard
signal outputs
mA output 2
setting ?
Variable allocation
Output range
Range measured value
Range control variable
Range correction value
22
mA output 1
measured value
0...20 mA
mA output 2
measured value
0...20 mA
Possible values
Initial value
Increment
as per identity
Measured value
code
Control variable
Correction value
0...20 mA
0...20 mA
4...20 mA
3.6/4-20 mA
0...1 ppm
0.01 ppm
-100 %...0 %
1 %
0...100 °C
0.1 °C
Control with standard signal
control
regulated value:
positive ClO
2
negative De-ClO
2
mA output 1
measured value
0 mA = 0.00 ppm
20 mA = 1.00 ppm
mA output 1
correcting value
0 mA =
0.0 °C
20 mA = 100.0 °C
Control with standard signal
control
regulated value:
positive ClO
2
negative De-ClO
2
mA output 2
measured value
0 mA = 0.00 ppm
20 mA = 1.00 ppm
mA output 2
correcting value
0 mA =
0.0 °C
20 mA = 100.0 °C
Lower value
Upper value
0 ppm
20 ppm
-100 %
+100 %
0 °C
100 °C
mA output 1
regulated value
0 mA =
0 %
20 mA = -100 %
D1C2-ClO2-027-GB
mA output 2
regulated value
0 mA =
0 %
20 mA = -100 %
D1C2-ClO2-032-GB
Remarks
If control is present, only
with adjustment variable
Reduction to 3.6 mA when
alarm relay switches
(not limit value violation)
Minimum range 0.1 ppm
Minimum range 1 %
Minimum range 1 °C