Layout - ProMinent DULCOMETER D1C Operating Instructions Manual

Measured variable chlorine dioxide
Table of Contents

Advertisement

7
Restricted Operating Menu / Layout
The restricted operating menu permits simple operation of the most important parameters. The following
overview shows the settings which can be selected:
ppm
0.60
auto.:
30.0 °C
mea. val 0.60 ppm
feed fwd:
70 %
ctrl out:
59 %
w:
0.60 ppm
calibration ClO
2
zero p.:
4.00 mA
slope:
6.50 mA/ppm
limits
setting ?
Only with control
control
setting ?
Setting in
complete
operating
menu
general setting
information
8
Permanent display 1
Permanent display 2
only with control
(w = setpoint)
calibration ClO
2
DPD-value:
0.60 ppm
temp.: 30.0 °C
limit 2
upper
1.50 ppm
limit 1
lower
0.10 ppm
control
control output
positive ClO
2
negative De-ClO
2
control
normal
control output
value:
30 %
or
control
with dead zone
control output
value:
30 %
or
control
manual
control output
value:
30 %
identcode: D1CA
DxDxxxxxxxxxx
software version
D1C-B1 FW-5.00
Positive values of setting variable: chlorine dioxide
Negative values of setting variable: De-ClO
calibration ClO
2
zero p.:
4.00 mA
slope:
6.75 mA/ppm
Access to setting menus can be blocked
For normal control
set point
0.60 ppm
Control with dead zone
set point 2 upper
0.80 ppm
set point 1 lower
0.60 ppm
For manual control
manual dosing
15 %
regulated range
alarm relay
active
2
(chlorine dioxide
destruction)
Number and scope of
setting menus depends
on the device.
with access code.
Proportional control
ctrl parameter
xp =
10 %
PID Control
ctrl parameter
xp =
10 %
Ti =
off
Td =
off
access c.:
5000
operating menu:
=
english
=
reduced
D1C2-ClO2-010-GB