ProMinent DULCOMETER D1C Operating Instructions Manual page 11

Measured variable chlorine dioxide
Table of Contents

Advertisement

Restricted Operating Menu / Description
Setpoint
Control parameter xp
Control parameter Ti
Control parameter Td
Manual metering
Abbreviations for control variables:
x
= 100 %/K
(inverse proportional coefficient)
p
p
T
= I controller integration time [s]
i
T
= D controller differential time [s]
d
General Settings
Access to all setting menus can be blocked with an access code !
general setting
information
Alarm relay
Access code
Language
Operating menu
Access Code
Access to the setting menu can be prevented by setting up an access code. The DULCOMETER
controller is supplied with the access code 5000 which permits free access to the setting menu. The
calibration menu remains freely accessible even if access to the setting menu is blocked by the code.
Possible values
Initial value
Increment
0.60 ppm
0.01 ppm
10 %
1 %
off
1 s
off
1 s
0 %
1 %
identcode: D1CA
DxDxxxxxxxxxx
software version
D1C-B1 FW-5.00
Possible values
Initial value
Increment
active
active
not active
5000
as per identity
as per identity
code
code
restricted
restricted
complete
Lower value
lower limit
measuring range measuring range
1 %
1 s
1 s
-100 %
alarm relay
active
Lower value
1
1
Upper value
upper limit
2 setpoints necessary
for control with dead zone.
Setpoint 1 < setpoint 2
500 %
xp referred to measuring range
9999 s
Function off = 0 s
2500 s
Function off = 0 s
+100 %
access c.:
operating menu:
=
english
=
reduced
Upper value
9999
Remarks
5000
D1C2-ClO2-016-GB
Remarks
®
D1C
11