Totalizers; A.9.6.3 Totalizers - Siemens SITRANS F Operating Instructions Manual

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Modbus communication
A.9 Modbus holding register tables
A.9.6.3

Totalizers

Totalizer 1 settings
Table A-30
Totalizer 1 settings
Modbus
Data type /
register
Size in bytes
8305
Unsigned / 2
8302
Unsigned / 2
8323
Float / 4
8325
Float / 4
8327
Float / 4
8329
Float / 4
8331
Float / 4
8306
Float / 4
112
Parameter
Description
Process value
Select process value
for totalization.
Fail-safe behav‐
Behavior of the totaliz‐
ior
er during the occur‐
rence of bad input val‐
ue.
Volume upper
Exceeding this limit
alarm limit
causes an alarm when
volume flow is selected
as input.
Volume upper
Exceeding this limit
warning limit
causes a warning when
volume flow is selected
as input.
Volume lower
Falling below this limit
warning limit
causes a warning when
volume flow is selected
as input.
Volume lower
Falling below this limit
alarm limit
causes an alarm when
volume flow is selected
as input.
Volume alarm
Hysteresis that affects
hysteresis
the alarm and warning
limits when volume
flow is selected as in‐
put.
Mass upper
Exceeding this limit
alarm limit
causes an alarm when
mass flow is selected
as input.
Default value
Value range /
[units]
Setting options
(units register)
0
● 0: Cut-off filtered
mass flow
● 1: Cut-off filtered
volume flow
0
● 0: RUN; totalization is
continued using the
bad input value.
● 1: HOLD; totalization
is stopped.
● 2: MEMORY;
totalization is
continued based on
the last incoming
good value.
Max. floating
point value
[m
]
3
(8321)
Max. floating
point value
[m
]
3
(8321)
Min. floating
point value
[m
]
3
(8321)
Min. floating
point value
[m
]
3
(8321)
0.0
-
[m
]
3
(8321)
Max. floating
point value
[kg]
(8320)
Operating Instructions, 09/2017, A5E41425845-AB
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
FST020 IP65 NEMA 4X

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fst020 ip65 nema 4x

Table of Contents