Path 1 settings
Table A-20
Path 1 settings
Modbus
Data type /
register
Size in bytes
1450
Float / 4
1458
Unsigned / 2
1465
Float / 4
User calibration
Table A-21
User calibration
Modbus
Data type /
register
Size in bytes
1000
Float / 4
1006
Float / 4
1004
Float / 4
FST020 IP65 NEMA 4X
Operating Instructions, 09/2017, A5E41425845-AB
Parameter
Description
Sensor frequen‐
Desired sensor frequency for
cy
path 1.
Path geometry
The number of crossings the sig‐
nal makes through the pipe (Re‐
flect or direct geometry for path
1).
E.g.:
Z path = 1
V path = 2
W path = 4
Spacing offset
Spacing offset (LTN) between
clamp-on sensors for path 1.
Parameter
Description
Slope correc‐
Single point flow correc‐
tion factor
tion applies to any flow
rate, positive or nega‐
tive.
Path 1 offset
User calibrated zero off‐
set for path 1.
Calibrated delta time =
delta time + Path 1 fac‐
tory offset + Path 1 off‐
set.
Path 1 factory
Factory calibrated zero
offset
offset for path 1.
A.9 Modbus holding register tables
Default value
Value range /
[units]
Setting options
(units register)
1100000
80000.0 -
[Hz]
4000000.0
1
1 - 10
0.0
-100.0 - 100.0
[m]
(8556)
Default value
Value range /
[units]
Setting options
(units register)
1.0
0.5 - 2.0
0.0
-1.25E-6 - +1.25E-6
[s]
0.0
-
[s]
Modbus communication
Access
level
Read /
write
Read /
write
Read /
write
Access
level
Read /
write
Read /
write
Read
only
99