Transmitter
7.4.6 Input signals
Active current
An individual analog power input is available for external analog devices.
input lin
The active current input lin is provided for connecting a two-wire transmitter with an out-
put signal of 4 – 20 mA.
Nominal input current
Maximum input current range
Internal power supply
Internal load resistance Rotamass
ROTAMASS
24 V
0 V
Fig. 36: Connection of external device with passive current output
①
Passive current
The passive current input lin is provided for connecting a four-wire transmitter with an out-
input lin
put signal of 4 – 20 mA.
Nominal input current
Maximum input current range
Maximum input voltage
Internal load resistance Rotamass
ROTAMASS
Fig. 37: Connection of external device with active current output
①
Iin+
Iin-
External passive current output device
Iin+
Iin-
External active current output device
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
Quick Reference Instruction Manual
Value
4 – 20 mA
2.4 – 21.6 mA
24 V
±20 %
DC
≤ 160 Ω
1
Value
4 – 20 mA
2.4 – 21.6 mA
≤ 32 V
DC
≤ 160 Ω
1
Wiring
45 / 54