Transmitter
Active status
Since this is a transistor contact, maximum allowed current as well as polarity and level of
output P/Sout
output voltage must be observed during wiring.
Load resistance
Internal power supply
ROTAMASS
24 V
0 V
Fig. 31: Active status output connection P/Sout
Active status output
P/Sout with internal
Internal pull-up resistor
pull-up resistor
Internal power supply
ROTAMASS
24 V
0 V
Fig. 32: Active status output P/Sout with internal pull-up resistor
P/Sout+
P/Sout-
External device with load resistance
P/Sout+
P/Sout-
External device
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
Quick Reference Instruction Manual
Value
> 1 kΩ
24 V
±20 %
DC
1
Value
2.2 kΩ
24 V
±20 %
DC
1
Wiring
43 / 54