Transmitter installation

6.4 Transmitter installation

6.4.1 Integral type

Rotating transmitter housing (integral type)
The transmitter housing can be installed in any one of four orientations.
Short-circuit hazard caused by penetrating water
WARNING
Failure of measuring electronics
▶ In order to prevent any water from penetrating the flow meter by way of the cable,
Insufficient sensor grounding connection
WARNING
Electric shock and ignition in hazardous areas
▶ Use a minimum torque of 4.3 Nm when tightening the screws.
Damage to flow meter
NOTICE
Rotating the transmitter housing several times in the same direction may damage the
connection between sensor and transmitter.
▶ Do not turn transmitter housing more than 270° in the same direction.
1. By using an Allen wrench (size: 5.0), remove the four fixing screws.
2. Lift transmitter housing.
install the transmitter in a way so that the cable gland is not pointed upward.
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
Quick Reference Instruction Manual
Installation
21 / 54