Transmitter
Active pulse
Connection of an electronic counter
output P/Sout
Maximum voltage and correct polarity must be observed for wiring.
Load resistance
Internal power supply
Maximum pulse rate
Frequency range
ROTAMASS
24 V
0 V
Fig. 26: Active pulse output connection P/Sout
Connection of an electromechanical counter
Maximum current
Average current
Internal power supply
Maximum pulse rate
Pulse width
ROTAMASS
24 V
0 V
Fig. 27: Active pulse output P/Sout connection with electromechanical counter
P/Sout+
1
P/Sout-
Load resistance
Electronic counter
P/Sout+
1
P/Sout-
Protective diode
Electromechanical counter
IM01U10A00-00EN-R, 2nd edition, 2017-08-29
Quick Reference Instruction Manual
Value
> 1 kΩ
24 V
±20 %
DC
10000 pulses/s
0 – 12.5 kHz
2
Value
150 mA
≤ 30 mA
24 V
±20 %
DC
2 pulses/s
20, 33, 50, 100 ms
2
Wiring