7.4
Sensor characteristics
The sensor characteristics menu shows:
● If a SENSORPROM
● Suppress error P 40 (SENSORPROM
● Sensor size
● Calibration factor
● Correction factor
● Excitation
See Sensor characteristics (Page 83).
Note
If a SENSORPROM is not installed, check the sensor characteristics to see if they match the
product label and the previous customer settings.
7.5
Reset mode
The reset mode is used for resetting/presetting totalizers or for restoring the factory settings.
Resetting
1. Press top up key
2. Enter password.
3. Use forward key
4. Press lock/unlock key
5. Press forward key
6. Press lock/unlock key
If restoring of factory settings (Page 89) is required:
1. Press lock/unlock key
See also reset menu diagram (Page 84).
7.6
Service mode
All outputs of the transmitter can be forced-controlled in the service mode menu, see Service
mode (Page 85).
It is possible to check whether the outputs are functioning.
Error pending and status log lists are also accessible from this menu and the operating time
(in days) can be read.
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018, A5E02083319-AH
is installed or not
®
not installed)
®
for 2 sec.
or backward key
to reach reset mode menu.
to enter reset menu.
to reach totalizer to be reset or default setting menu.
to start resetting.
again to confirm destruction of customized settings.
Functions
7.6 Service mode
51