Alarm, error, and system messages
8.1 Diagnostics
Acceptance levels
The following three acceptance levels are selectable.
● Fatal error: Only fatal errors are registered as errors
● Permanent error: Permanent and fatal errors are registered as errors
● Warning (Default value): Warnings, permanent and fatal errors are registered as errors
Error information is displayed in title and subtitle lines, see display layout (Page 41). Title line
will show time in days, hours and minutes since occurrence of error. Subtitle line will flash
between an error text and a remedy text. Error text will indicate type of error (I, W, P or F),
error number, and error text. Remedy text will inform operator of action to take to remove error.
Status log
The latest 9 errors are stored in the status log. Errors are stored in the status log for 180 days,
even if they are corrected.
Alarm field
The alarm field on the display will always flash when an error is pending.
Error output
The digital and relay output can be activated individually error by error (error level). The relay
output is default selected to error level. An output can also be selected to activate on a single
error number.
The alarm field, error output and error pending always operate together.
Operator menu
Error pending and status log are as default enabled ( ) in the operator menu.
54
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018, A5E02083319-AH