Commissioning
6.1 Local user interface
Figure 6-2
Note
Isolate the device from power when removing the glass lid
If the glass lid is rotated during operation, the capacitive touch keypad automatically calibrates
to align the distance from lid to keypad. To avoid any unintended programming during this
calibration procedure, it is recommended to always isolate the device from power before
removing the glass lid and until it is properly fixed to the device again.
Mode field symbols
Lock field symbols
Keypad
The keypad is used to set the flowmeter. The keys function as follows:
TOP UP KEY
FORWARD KEY
42
Capacitive touch keypad
Communication mode
Service mode
Operator menu
Product identity
Ready for change
Value locked
This key (when held for 2 sec.) is used to switch between op‐
erator menu and setup menu. In transmitter setup menu, a short
press will cause a return to previous level.
This key is used to step forward through the menus. It is the
only key normally used by the operator.
Language mode
Basic settings
Output
External input
Access to submenu
RESET MODE: Zero setting of totalizers and initializa‐
tion of setting
Operating Instructions, 08/2018, A5E02083319-AH
Sensor characteristics
Reset mode
Operator-active
Operator-inactive
SITRANS MAG 6000 I/6000 I Ex de