Siemens SITRANS MAG 6000 I Operating Instructions Manual page 15

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS MAG 6000 I:
Table of Contents

Advertisement

Sensor coil terminals
85 and 86
Power supply:
Parameter
Supply
Max. allowable supply voltage U
ATEX certificate)
Ambient temperature
Enclosure
Fail-safe protection
With intrinsically safe circuits, use only certified meters appropriate for the transmitter.
If a non-conforming suuply unit is used, the "fail-safe" type of protection wil no longer be
effective and the approval certification will be invalid.
User I/O interface
Remote MAG 6000 I: Model 7ME693-2BA4/5
Compact:
MAG 6000 I with MAG 3100 series: Model 7ME63x0-xxxxx-xD/E
Passive current
Active current
(31, 32)
(31, 32)
IIC
U
28 V DC
U
i
O
I
100 mA
I
I
O
C
19.7 nF
C
I
O
L
36 µH
L
I
O
P
0.7 W
P
I
O
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018, A5E02083319-AH
WARNING
IIB
IIC
30 V DC
30 V DC
87.8 mA
87.8 mA
557 nF
63 nF
18.4 mH
4.6 mH
0.66 W
0.61 W
Terminal type
"e"
MAG 6000 I Ex de
115 to 230 VAC or 18 to 30 VDC
(according to
264 V
m
-20 to +60 °C (-4 to 140 °F)
IP67 / NEMA 4X
Relay
Frequency/pulse
(44, 45, 46)
(56, 57)
IIC
IIC
U
30 V DC
U
28 V DC
i
i
I
200 mA
I
100 mA
I
I
C
3.3 nF
C
14.2 nF
I
I
L
0 µH
L
36 µH
I
I
P
1.2 W
P
1.2 W
I
I
Safety notes
2.3 Installation in hazardous area
Profi (FISCO)
Digital input
(95, 96)
(77, 78)
IIC
U
17.5 V DC
U
i
i
I
380 mA
I
I
I
C
0 nF
C
I
I
L
0 µH
L
I
I
P
5.32 W
P
I
I
IIC
30 V DC
0 nF
0 µH
1.2 W
15