3.14.2 Operating notes
The lamp can be opened and closed directly on the CLH 606 A.
will light up yellow if the lamp is on and the light from the halogen lamp is
available.
Note
The spectral emission radiation characteristic of the halogen bulb
will change during the warm-up phase. A minimum warm-up time
of approx. 10 minutes must therefore be observed as the emitter is
then able to achieve a stable operating condition. For very precise
measurements, a warm-up time of 60 minutes is necessary so that the
optical and mechanical components of the bulb can achieve a constant
working temperature.
3.14.3 Safety information for CLH 606 A lamp cassette
Caution
Coherent radiation!
Never look directly into the measuring beam. Coherent radiation can
lead to eye injuries (e.g. conjunctivitis). Please observe the relevant
regulations when handling light sources of this type.
3.14.4 Spare parts for CLH 606 A lamp cassette
Designation
Halogen lamp 10V/1.8A for CLH 606
4
Cat. no.
000000-1446-136