3.2.7 Connecting the system
1. Connect the power cable on the rear side of the housing
into the power socket.
2. Connect your PC's ethernet cable to the ethernet port on the front side of
the housing.
4
3. The 37-pin D-SUB port for 16 x IN/OUT (for industrial use, shielded)
the 32-pin port for 16 x IN/OUT (for laboratory use, unshielded)
signed to establish connections with external digital interfaces (e.g. external
triggers, remote controls).
4. Turn on your PC and launch the program.
5. Turn on the system. The LED
ready to be used.
3.2.8 Replacing slide-in cassettes
Please ensure that the system is disconnected before inserting/removing slide-in
cassettes.
1. On the ECCU
, set the switch
3
off, then disconnect the device from the power supply and remove the cable
and/or light guide.
2. Loosen the securing screws on the front of the cassette and pull out the
cassette.
3. Place the new cassette on the module support and slide it into the housing
until you feel it snap into place.
4. Tighten the cassette firmly with the securing screws and reconnect the cables
and/or light guides.
5. Plug in the power cable and turn the device on.
will illuminate green when the device is