3.4.2 Operating notes
Six LEDs are installed on the front of the MCS 611 NIR 2.2 spectrometer cassette.
During operation, the specified LED will light up green (BUSY)
rupting the saving of spectra (with short integration times, the LED will light up
constantly). The LED labeled TRIGGER
if the measurement is triggered by the hardware. The LED labeled POWER
light up green if the cassette is switched on. In the event of a cassette error, the
red LED labeled ERROR will light up.
on, first the yellow LED labeled SETTLING
the cassette is ready for measuring. Once the temperature LED turns off (orange
LED labeled TEMP
), the spectrometer is at the configured temperature and is
9
ready for measuring. If the LED does not turn off, the cooling is defective, and the
slide-in cassette must be replaced. If the LEDs do not light up as intended, a failu-
re has occurred, and the slide-in cassette must be replaced.
3.4.3 Changing the spectrometer cassette
Please ensure that the system is disconnected before inserting/removing slide-in
cassettes.
1. On the ECCU, set the switch to the position in which all LEDs are off, then
disconnect the device from the power supply and remove the cable and/or
light guide.
2. Loosen the fastening screws
cassette.
3. Place the new cassette on the module support and slide it into the housing
until you feel it snap into place.
4. Tighten the cassette firmly with the securing screws and reconnect the cable
and/or light guide.
5. Plug in the power cable and turn the device on.
Note
Submit defective slide-in cassettes to our Customer Service team for
repair. The address can be found on page 2 of this document.