3.8.2 Safety information for BLX 600/SMA xenon lamp
Warning
The BLX 600/SMA xenon lamp contains electronic, electromechanical,
mechanical and optical components. These components require special
care when being transported and used. The ventilation slots on the rear
side of the housing must not be covered or obstructed.
Caution
Risk of explosion!
The information and safety guidelines provided by the lamp
manufacturer must be consulted before changing the lamp. The xenon
lamp's gas filling is highly pressurized even when the lamp is cold (10
bar at room temperature). The lamp must be allowed to cool for at
least 30 minutes before being changed. For your own safety, wear face
protection, a leather apron and cuffed leather gloves as protection
against shattering. Be sure to dispose of the used lamps appropriately.
Caution
Coherent radiation!
The xenon lamp emits light in the ultraviolet range! Do not look directly
into the beam path or expose your skin to it! Doing so will result in
injury to the eyes and skin. Please observe the relevant regulations
when handling UV light sources. Cover unused SMA connectors on the
lamp with caps.
Warning
The xenon lamp's ignition voltages are in the range of several 100 V.
These voltages are generated in and accumulate in the electronics. Do
not attempt to modify the internal electronics or wiring!
3.8.3 Spare parts for the BLX 600/SMA lamp cassette
Designation
Flash Bulb L4633-01
Cat. no.
000000-0235-825