Calibrating Ohms Measure

Connect the UUT as shown in Figure 32.
After making the proper connections, type "9" then press the Enter key on the PC
keyboard. Input directions are displayed at the bottom of the calibration menu:
Enter Selection:
Input 15 Ohms, Press Enter When Stable
Input 400 Ohms, Press Enter When Stable
Input 4000 Ohms, Press Enter When Stable
Use the 5520A to input 15 Ω, 400 Ω, and finally 4000 Ω. The Enter key must be applied
prior to advancing between steps.