Cal ISO mA

Connect the UUT as shown in Figure 23.
From the Calibration Menu, enter the cal step for Cal ISO mA. The PC displays:
Enter 0 ma
Set the Fluke 5520A to output 0.0000 mA. After a short while, the PC displays the
following calibration constants and prompt:
Offset =
Enter 24.00 ma
Set the Fluke 5520A to output 24.0000 mA then press the space bar on the PC. After a
short while, the PC displays the following calibration constants:
diff = (Counts - Offset)
32133
=
mA per count =
- press space bar to continue
Press the space bar to return to the Calibration Menu.
Adjustment is completed. Turn the UUT off, then back on, to return to normal operation.
Figure 23. ISO mA Calibration Connections
- press space bar to continue
-323
- press space bar to continue
31810
-
-323
0.000747
Calibrators
Calibration Adjustment
aal05f.eps
53