Pressure Module Input (725/726); Calibration Adjustment; Setup - Fluke 724 Calibration Manual

Hide thumbs Also See for 724:
Table of Contents

Advertisement

Pressure Module Input (725/726)

1. Connect a Fluke 700 Series Pressure Module to the 5-pin LEMO connector at the top
of the UUT; then press U.
2. Verify that the display first shows -----psi, then changes to a pressure value.
3. Disconnect the pressure module from the UUT.
The performance tests for the 725 and 726 are now complete. Disconnect and secure all test
equipment.

Calibration Adjustment

The Calibrators have electronic calibration. There are no mechanical adjustments and the
calibration is done case closed. The calibration is done via serial communications port, by
sending and receiving commands and readings. For the 725 and 726 serial port
connection, you may use the 700SC Serial Interface Cable, PN 667425. This will permit
communication through the pressure port connection. The 724/725 Calibration Cable, PN
1556747 is required for the 724
the 10-pin header in the battery compartment.
Throughout these procedures, the Calibrators are referred to as "UUT" (Unit Under Test).
Two methods of calibration are available for the UUT: using a serial-based program via a
PC (see Table 6 for requirements), and using a Met/Cal calibration procedure. This
manual only describes the serial-based PC program. The automated Met/Cal procedures
are available from the Fluke Metrology Group. For more information on the automated
Met/Cal procedures contact Fluke or visit the Fluke Web site at www.fluke.com.
As long as the calibrators have been within at a stable temperature within the range of
º
º
18
C to 28
C for an hour or more, they only need 5 minutes to warm up. If temperature
conditions were previously below 10
allowed to stabilize for a minimum of 3 hours prior to calibration.
The Calibrators should be calibrated at the required interval (defined by the factory as 1
year).

Setup

Setup the PC and UUT as follows:
1. Ensure that the terminal communications software is installed on the PC (such as
Hyper Terminal or PcPlus).
2. Connect the interface cable to the appropriate connector on the UUT. Remove the
jumper beside the ten-pin connector when using the 724/725 Calibration Cable on a
725. Replace the jumper after calibration is complete.
3. Connect the 9-pin 'D' connector to the PC serial port. An adapter may be needed for
PCs that use 25-pin 'D' serial connectors.
4. Verify the settings on the PC are as follows:
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None
Local echo on
, and can be used with the 725 and 726.
º
C or higher then 40
Calibration Adjustment
. This connects to
º
C, then the Calibrator must be
Calibrators
39