724/725/726
Calibration Manual
Cal mV Output
Connect the UUT as shown in Figure 18.
From the Calibration Menu, enter the cal step for Cal mV Output. The PC displays:
Set the Fluke 8508A to read V DC. When the reading on the Fluke 8508A has stabilized,
enter the value in millivolts on the PC, then press Enter. You only need to enter four
places past the decimal point and do not need to enter the units (mV). After a short while,
the PC displays:
Enter the mV reading from the Fluke 8508A on the PC, press Enter, then press the space
bar. After a short while, the PC displays the following calibration constants and new
prompt:
6537.086487
87
- press space bar to continue
Press the space bar to return to the Calibration Menu.
46
Figure 18. mV Output Calibration Connections
Zero into DAC. Enter mV displayed :
Max value into DAC. Enter mV value displayed :
=
111.874998
-
-12.450000 )/ 65279.00 mV per count =
aal08f.eps
6537.0864