Lower Display Frequency Measurement Test (725 And 726) - Fluke 724 Calibration Manual

Hide thumbs Also See for 724:
Table of Contents

Advertisement

724/725/726
Calibration Manual

Lower Display Frequency Measurement Test (725 and 726)

1. Set the 5520A to source a 10 kHz, 1 V peak-to-peak square wave (use the blue
2. Press F on the UUT (Kfor 726) until MEASURE and kHz appear on the
3. Make the connections shown in Figure 6.
4. Verify that the UUT frequency reads between 9.98 kHz and 10.02 kHz.
5. Press STBY on the 5520A.
30
softkey under the wave type to change the wave shape).
lower display.
Figure 6. Lower Display Frequency Test Connections
aal04f.eps