Calibrating Voltage Input

Connect the UUT as shown in Figure 28.
After making the proper connections, type "5" then press the Enter key on the PC
keyboard. Input directions are displayed at the bottom of the calibration menu:
Enter Selection:
Input 0 Volts, Press Enter When Stable
Input 20 Volts, Press Enter When Stable
Use the 5520A to input 0 V and then 20 V.
Figure 28. Voltage Input Calibration Connections
Calibrators
Calibration Adjustment
aal04.eps