724/725/726
Calibration Manual
Command
TC_TYPE
TC_TYPE?
TSENS_TYPE
TSENS_TYPE?
CJC_STATE
CJC_STATE?
RTD_TYPE
RTD_TYPE?
CPRT_R0
CPRT_R0?
CPRT_MIN_T
CPRT_MIN_T?
CPRT_MAX_T
20
Table 5. Serial Command List-726 (cont.)
Response/Actions
Returns TC type
B, C, E, J, K, L, N, R, S, T, U,
BP, XK, MV
Returns the sensor type TC or
RTD
Returns ON or OFF
Returns RTD type
NI120, PT392_100,
PT385_100, PTJIS_100,
PT385_200, PT385_500,
PT385_1000, CU_10, CUST1,
CUST2, CUST3, OHMS
Returns CPRT R0 with units
OHM.
Returns {number} CEL.
Command
Comment
Arguments
One argument, valid
Set TC type
settings: B, C, E, J, K,
L, N, R, S, T, U, BP,
XK, MV
Verify TC type
1 argument TC or
Sets the
RTD
sensor type
TC or RTD
Verify is set
for RTD or TC
One argument ON
Thermocoupl
or OFF
e cold
junction
compensation
Verify CJC
state
1 argument: NI120,
Sets RTD
PT392_100,
type
PT385_100,
PTJIS_100,
PT385_200,
PT385_500,
PT385_1000,
CU_10, CUST1,
CUST2, CUST3,
OHMS
Verify the
RTD type
setting
2 arguments
{number} OHM.
Sets the custom
CPRT R0.
RTD_TYPE
must be either
CUST1,
2 arguments
CUST2 or
{number} CEL.
CUST3
2 arguments
{number} CEL.
Ch 1
Ch 2
X
X
X
X
X
X
X
X
X
X
X
X
X

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

726725

Table of Contents