Calibrating Isolated mA Input
Connect the UUT as shown in Figure 24.
Figure 24. Isolated mA Input Calibration Connections
After the proper connections have been made, type "1" (the first calibration step) then
press the Enter key on the PC keyboard. Input directions are displayed at the bottom of
the calibration menu:
Enter Selection:
Input 0 mA, Press Enter When Stable
Input 24 mA, Press Enter When Stable
Use the 5520A to input 0 mA and then 24 mA.
Calibrators
Calibration Adjustment
aal05f.eps
55