Lower Display Ohms Source Tests - Fluke 724 Calibration Manual

Hide thumbs Also See for 724:
Table of Contents

Advertisement

724/725/726
Calibration Manual

Lower Display Ohms Source Tests

1. Press F ( R on the 726) on the UUT until Ω appears on the lower display. If
2. Set the Fluke 8508A to measure 4-wire resistance.
3. Set the Fluke 8508A to 2 k range for 72x resistance < 400 Ω, to 20 kΩ range for
4. Make the connections shown in Figure 12.
5. Use the arrow keys on the UUT to set the UUT output to the resistance values in
UUT Settings
The performance tests for the 724 are now complete. Disconnect and secure all test
equipment.
38
necessary, press M until SOURCE appears on the lower display.
400 Ω or more.
Table 18 and verify that the Fluke 8508A readings are within the limits shown.
Figure 12. Lower Display Ohms Source Test Connections
Table 18. Lower Display Ohms Source Readings
Fluke 8508A Readings
15.0 Ω
14.9 Ω to 15.1 Ω
360.0 Ω
359.9 Ω to 360.1 Ω
500 Ω
499.5 Ω to 500.5 Ω
1500 Ω
1499.5 Ω to 1500.5 Ω
3200 Ω
3199.0 Ω to 3201.0 Ω
3800 Ω
724/725
Fluke 8508A Readings
14.94775 Ω to 15.05225 Ω
359.896 Ω to 360.104 Ω
499.375 Ω to 500.625 Ω
1499.275 Ω to 1500.725 Ω
3798.93 Ω to 3807.01 Ω
-
aal13f.eps
726
Range
2 kΩ
2 kΩ
20 kΩ
20 kΩ
-
20 kΩ
20 kΩ