Calibrating mA Input
Connect the UUT as shown in Figure 26.
After making the proper connections, type "3" then press the Enter key on the PC
keyboard. Input directions are displayed at the bottom of the calibration menu:
Enter Selection:
Input 0 mA, Press Enter When Stable
Input 24 mA, Press Enter When Stable
Use the 5520A to input 0 mA and then 24 mA.
Figure 26. mA Input Calibration Connections
Calibrators
Calibration Adjustment
aal06f.eps
57