Cal mA Input (Fluke 725 Only)

Connect the UUT as shown in Figure 15.
From the Calibration Menu, type the cal step for Cal mA Input. The PC displays:
Enter 0 ma
Set the Fluke 5520A to output 0.000 mA, let the reading settle a few seconds, then press
the space bar on the PC. After a short while, the PC displays the following calibration
constant and new prompt:
Offset =
Enter 24.00 ma
Set the Fluke 5520A to output 24.000 mA, let the reading settle a few seconds, then press
the space bar on the PC. After a short while, the UUT displays the following calibration
constants and new prompt:
diff = (Counts - Offset)
8106924
mA per count =
press space bar to continue
Press the space bar to return to the Calibration Menu.
Figure 15. mA Input Calibration Connections
- press space bar to continue
-337
- press space bar to continue
=
8106587
-
-337
0.000003
Calibrators
Calibration Adjustment
aal06f.eps
43