3
Chapter 3 Configure the instrument and manage instrument profiles
Learn about the instrument
Restore factory
defaults
Select a language
Learn about the instrument
View instrument
specifications
34
IMPORTANT! Back up the instrument before restoring factory defaults (see "Backup
or restore the instrument" on page 70).
1. In the home screen, touch
Defaults.
2. Touch Restore Factory Defaults.
3. Power Off, then power On the instrument to apply the change.
• Instrument profiles and files stored on the instrument are deleted, including
all EDS files and any custom dye calibrations.
• The most recent valid ROI/Uniformity, Background, and system dye
calibrations are not deleted.
After restoring factory defaults, perform initial start-up tasks. See "Initial start-up" on
page 25.
1. In the home screen, touch
2. Touch the desired language selection.
3. Touch OK.
In the home screen, touch
information.
"View instrument specifications" on page 34
"View and export the License Agreement" on page 35
In the home screen, touch
access the following information.
• Model Name
• Wired IP Address
• Wireless IP Address
• Instrument Serial Number
• Block Serial Number
• UUID
• Firmware Version
• Thermal Block Controller
• Instrument Server
• User Interface
QuantStudio
Settings4Instrument Settings4Restore Factory
Settings4Instrument Settings4Language.
Settings4About Instrument to access the following
Settings4About Instrument4About Instrument to
1 Real-Time PCR System Installation, Use, and Maintenance Guide
™