Reset; Basic Settings; Ph Measurement - wtw inoLab pH/Cond Level 1 Operating Manual

Measuring instrument for ph, redox, and conductivity
Hide thumbs Also See for inoLab pH/Cond Level 1:
Table of Contents

Advertisement

Operation

Basic settings

pH measurement

Conductivity mea-
surement
118
4.8

Reset

You can reset (initialize) the settings for the pH and conduc-
tivity measurement separately.
The following functions for the pH measurement (pH InI) are
reset to the values they had on delivery:
Measuring mode
Asymmetry
Slope
Calibration procedure
Temperature, manual
The following functions for the conductivity measurement
(
InI) are reset to the values they had on delivery:
Measuring mode
Cell constant
Temperature compensation
Reference temperature
Temperature coefficient of
the linear temperature com-
pensation
TDS factor
Note
After a reset, the calibration data are lost. Thus, calibrate af-
ter a reset.
pH
0 mV
-59.16 mV/pH
AutoCal TEC
25 °C
-1
0.475 cm
(calibrated)
-1
0.475 cm
(set up)
nLF
2.000 %/K
1.00