Setting Up The Temperature Compensation Tc - wtw inoLab pH/Cond Level 1 Operating Manual

Measuring instrument for ph, redox, and conductivity
Hide thumbs Also See for inoLab pH/Cond Level 1:
Table of Contents

Advertisement

ips
Application
notes
4.6

Setting up the temperature compensation TC

The calculation of the temperature compensation is based
on the preset reference temperature, Tref 20 or Tref 25 (see
section 4.7 C
ONFIGURATION
You can select one of the following temperature compensa-
tions:
l
Non-linear temperature compensation "nLF"
according to DIN 38404 or EN 27 888
l
linear temperature compensation "Lin" with a coeffi-
cient that can be set in the range 0.001 ... 3.000 %
l
no temperature compensation
Note
Select the following temperature compensations to work
with the test samples given in the table:
Test sample
Natural water
(ground water,
surface water,
drinking water)
Ultrapure water
Other aqueous
solutions
Salinity
(seawater)
).
Temperature
compensation TC
nLF
according to DIN 38404
EN 27 888
nLF
according to DIN 38404
EN 27 888
Set linear
temperature coefficient
0.001 ... 3.000 %/K
Automatically nLF accord-
ing to IOT
Operation
/
K
Display
indicator
Sal,
111

Hide quick links: