2.1.8 Flanged style installation
Figure 2-7: Flanged style installation
A. Installation bolts and nuts (supplied by customer)
B. Gaskets (supplied by customer)
C. Flow direction
Note
The required bolt load for sealing the gasket joint is affected by several
factors, including operating pressure, gasket material, width, and condition. A
number of factors also affect the actual bolt load resulting from a measured
torque, including condition of bolt threads, friction between the nut head and
the flange, and parallelism of the flanges. Due to these application-dependent
factors, the required torque for each application may be different. Follow the
guidelines outlined in ASME PCC-1 for proper bolt tightening. Make sure the
flowmeter is centered between flanges of the same nominal size as the
flowmeter.
2.1.9 Insert integral temperature sensor (MTA option only)
About this task
Note
Step number of procedure corresponds with number in
Quick Start Guide
A
B