Environmental Conditions - Mcube CUBEscan BioCon-900S Operator's Manual

Table of Contents

Advertisement

Item
Use
Power
Output
Insulation
Water Resistance
Mode of Operation
Weight

ENVIRONMENTAL CONDITIONS

9.5
WARNING
Do not leave this device in places subject to extremely high temperatures such as a sealed vehicle
CAUTION
or in direct sunlight. This can cause a fire.
 Do not place the heavy object on the device. The device can be damaged.
Do not place the device on an unstable surface. This can cause the device to fall or tip over and
 Keep away from the humid or dusty place.
cause injury.
CAUTION
Do not place the heavy object on the device. The davice can be damaged.
Keep away from the humid or dusty place.
Use the device indoors only under the environmental conditions.
For additional protection of the device during a lightning storm, or when it is left unattended and unused
for long periods of time, unplug it from the wall outlet.
The immediate use of the probe which has been stored under 10℃ without warming may damage the
probe. So take it into a room where the temperature is over 10℃ and let BioCon-900S warm up for a
while. And then when the temperature of probe is over 10℃ you may use it.
If you are not using BioCon-900S more than 3 months, remove the battery to prevent it from
discharging and store it in accordance with the recommended conditions, and charge the battery every
6 months.
Contact your local distributor or Mcube Technology, when you need to remove or replace the battery.
Use
Ambient temperature range
Relative humidity
Atmospheric pressure range
Ambient temperature range
Features
Indoor
AC/DC Adapter:
Manufacturer: MEAN WELL
Model: GSM60A12
Input: 100- 240VAC, 50/60Hz, 1.4-0.7A
Output: 12V
Wireless Power Transfer Charger
Class I with protective earth
IPX0
Continuous operation.
480g (without adapter and AC cord)
Enviromental Specifications
Operating conditions
Storage and transport conditions
Use
TM
CUBEScan
BioCon-900S Operator's Manual
Charger
5.0 A, 60W MAX
+10 - +40℃ (+50 - +104℉)
+30% - +75% non-condensing
+700hPa - +1060hPa
-20 - +60℃ (+14 - +140℉)
Indoor
Indoor
Page 56