WARNING
 "LASER RADIATION – AVOID DIRECT EXPOSURE TO BEAM"
 Do not capture human being with the barcode module.
 Hold the probe with both hands tightly, when you read barcodes with patient information.
 Recording barcodes is not available, after the scan result is saved.
Figure 22 Information Screen
How to read a barcode
1) Tap the Barcode icon(
2) Move the aiming position to the center of the barcode and press the Scan Button
3) A red light guide is emitted.
4) Keep pressing the Scan Button
5) The barcode reader decodes continuously for a good barcode scan.
6) You have a good barcode with a short beep. The barcode numbers is displayed on Patient Tag of the
Information screen.
<good>

MEASURE BLADDER VOLUME

5.5
NORMAL SCAN
5.5.1
The normal scan is executed only when Pre-scan setup is 'Off(
"7.7 PRE-SCAN SETTING"
1) The patient should be lying in a supine position and lift up the patient garment exposing the abdominal
region from the pubis to the navel.
2) Place an ample amount of gel without air bubbles to the probe head. In order to assure optimal
transmission of energy between the patient and the probe, a conductive gel must be applied on the
probe head.
3) Place the probe head on the patient's abdomen with the screen facing upwards and aim it to the location
where the bladder is expected to be above 3~4cm the pubic bone.
4) Ensure that the probe head is pointing toward the patient's foot.
Figure 23 Barcode Scan Screen
) on the Information screen.
and do not release until getting barcodes.
<bad>
TM
CUBEScan
BioCon-900S Operator's Manual
)' on the Setup screen. See the section
with thumb.