Scan Type - Mcube CUBEscan BioCon-900S Operator's Manual

Table of Contents

Advertisement

Status icon (
) is always displayed in the Status at the top of the screen, indicating the current
capacity. Before using the probe for the first time, you must fully charge the battery. When you don't use
your probe, we recommend that you mount and store it on the Charger to ensure that the probe is
sufficiently charged. The battery is protected by protection circuit module and have the functions of
overcharge, overdischarge and overcurrent prevention to maintain safety and prevent significant
deterioration of cell performances.
To conserve battery life, the probe is configured to go into sleep mode. The probe goes into sleep mode
when you press the Power Button(
with power on. You can adjust the time for sleep mode in the setup(2min/3min/5min), if necessary. See the
section"7.4 SLEEP MODE SETTING". Shortly press the Power Button(
Figure 6 Probe-mounting
1) Connect the AC/DC adapter to the power cord.
2) Connect the AC/DC adapter to the Charger
3) Plug the AC/DC adapter into an electric outlet
4) The battery indicator (3) on the Charger lights on in steady green (Power on).
5) Mount the probe on the Charger. Make sure that the upper part of probe reaches the wall of Charger.
See Figure 5.
6) As soon as the probe is mounted on the Charger, the beep and light indicators work as below;
a) A short beep: A connection for the probe's battery charging is completed.
b) The battery indicator (3) on the Charger lights on in blinking green(Charging in progress).
c) The battery indicator (2) on the probe lights on in steady orange (Charging in progress).
7) The Power indicator (1) is turned on in steady yellow when the probe is ready to use.
8) It takes approximately 6 hours for the discharged battery to fully charge. When fully charged, the battery
indicator(2) on the probe turns off.

SCAN TYPE

4.3
You can adjust one of the following options to measure bladder volume:
Pre-scan: General 2D real-time ultrasound scanning. 2D ultrasound images are displayed continuously in
live image basis. This helps an operator locate bladder position and predicting the range of residual urine
to detect the bladder before normal scan.
Normal Scan: Get 12-plane ultrasound images and calculate the residual urine in the bladder (3D
scanning).
CUBEScan
) twice or if the probe has not been touched in the designated time
Figure 7 Probe Indicatior
TM
BioCon-900S Operator's Manual
) to wake up the system.
Figure 8 Charger Indicator
Page 19