Battery Maintenance; Weekly Inspection; Disposal; Device Repair - Mcube CUBEscan BioCon-900S Operator's Manual

Table of Contents

Advertisement

Disinfection
8.3.3
Please note the that low-level disinfection of the probe head is required before a new measurement.
a) Put on new gloves
b) Spray the probe head until it is sufficiently wet with a Metrex CaviCide spray.
Figure 75 Probe Head
c) Leave the solution on the probe head surface for about 3 minutes.
d) If you spray the Metrex CaviCide spray additionally, leave the solution on the surface for at least 3
minutes.
e) Allow the probe head to air dry.
Wipe the probe head thoroughly using the fresh CaviWipes.
f)
g) Keep the probe head wet for 3 minutes.
h) Additional wipes may be used to ensure the probe remaining wet for the 3 minutes, if necessary.
Allow the probe head to air dry.
i)
Using sterile, lint-free cloths wetted with RO/DI water thoroughly wipe the probe for at least a minute
j)
k) Allow the probe head to air dry.

BATTERY MAINTENANCE

8.4
BioCon-900S battery cannot be removed by the user. If you do not plan to use BioCon-900S for a long
period of time, you should switch the probe to shutdown mode to prevent its battery from fully discharging.
The BioCon-900S should be charged every 6 months at a minimum, store it in accordance with the
recommended conditions. See the section "9.5 ENVIRONMENTAL CONDITIONS".

WEEKLY INSPECTION

8.5
a) Inspect the probe and Charger thoroughly for cracks or leakages.
b) Inspect the power cord for checking if there are any damaged parts.
c) When scanning, check out any abnormal noise emanating from the probe head.

DISPOSAL

8.6
The device and components may contain environmentally hazardous materials (mineral oil, lead, battery
pack, etc.). When they have reached the end of its useful service life, return them to the Mcube Technology,
or follow your local regulations for hazardous waste disposal.

DEVICE REPAIR

8.7
Faults not described in section "8.1 TROUBLESHOOTING" are intended to be serviced by a certified
technician. When any troubles unlisted in section "8.1 TROUBLESHOOTING" occur, contact an authorized
service provider or Mcube Technology.
TM
CUBEScan
BioCon-900S Operator's Manual
Page 46