4. SETUP

See the section "4.1 ACCURACY OF MEASUREMENT" before use.
Check the probe's battery has sufficient power through battery icon. If the battery has not been
sufficiently charged, recharge the battery before bladder measurement.
Wipe the probe head gently with a soft cloth dampened in cleaning and disinfection agents to clean and
disinfect the probe. See the section "8.3 CLEANING & DISINFECTION".

ACCURACY OF MEASUREMENT

4.1
Following cases affect ultrasound transmission and accuracy of measurement.
When a patient has had supra-pubic or pelvic surgery.
A patient with catheter in his or her bladder.
A patient with scar, sutures, staples or incisions in his or her abdomen.
Use of unsuitable ultrasound gel .
Air bubbles between probe head and the skin of the patient.
An excessive obese patient.
Use of the probe with an unclean probe head.
Inappropriate patient position.

CHARGE THE BATTERY

4.2
WARNING
The battery must be replaced by the authorized service provider only.
Use only the provided Charger and power cord to charge the battery for electric safety.
Do not disassemble, heat or incinerate the battery.
Keep battery out of reach of children and in original sealed package.
Dispose of used batteries promptly according to local recycling or waste regulations.
To reduce the risk of leakage, explosion, fire, or serious injury, note the following when handling the
lithium-ion battery included in the system:
- Never short-circuit the battery by either accidentally or intentionally bringing the battery terminals
into contact with any other conductive objects. This could cause serious injury or fire and could also
damage the battery and BioCon-900S.
- Never expose the battery to abnormal shock, vibration or pressure. The battery's internal
protective covering could fail, causing it to overheat or ignite, resulting in caustic liquid leakage,
explosion or fire.
Avoid charging near fire or in the sunlight.
Do not use the Charger for other equipment.
CAUTION
The Charger, power cord, and adapter are not intended for patient contact. Ensure 6 ft (2 m) are
maintained between the patient and these components.
Immediately discontinue use of the battery, if the battery emits an unusual smell, feels hot, changes
color or shape, or appears abnormal in any other way. Contact Mcube Technology or your local
distributors.
The probe is powered by a internal lithium ion battery. The battery is charged by Charger and the Battery
TM
CUBEScan
BioCon-900S Operator's Manual
Page 18