SELF TEST

8.8
1) Put the probe on the Charger
2) Turn the probe on and tap the Setup icon (
3) Tap the Next icon (
4) Tap the Scan icon (
icon (
). Tap the Exit icon (
5) Self-test progress bar appears, when the calibration is started. The progress bar indicates the
progress of calibration. It takes about 3minutes to complete the calibration. When the calibration is
completed, 100% sign is displayed.
Tap the Ok icon (
the Setup screen.
) on the Setup screen and select the Self-test icon (
) to start the calibration. You can cease the Self-test by tapping the Cease
) to return to Setup screen, if you want.
) and the latest calibration date is updated. Tap the Exit icon (
TM
CUBEScan
BioCon-900S Operator's Manual
Self-test
) on the bottom right side of the Home screen.
)
) to return to
Page 47