Cleaning & Disinfection - Mcube CUBEscan BioCon-900S Operator's Manual

Table of Contents

Advertisement

CLEANING & DISINFECTION
8.3
WARNING
Cleaning is a necessary step to perform before disinfecting the device. If the device is not cleaned
properly, even after the disinfection procedure is completed, there may be contamination of the
device.
Incorrect cleaning or disinfecting of any part of the system can cause permanent damage.
Do not use solvents such as thinner or benzene, or abrasive cleaner on any part of the system.
Do not subject the system to any method of sterilization.
Do not immerse the probe except the probe head in cleaning or disinfectant solution.
Do not immerse the Charger in cleaning or disinfectant solution.
Do not use metal or abrasive brushes. These may scratch the instrument causing permanent device
damage.
Wear protective gloves when cleaning the device.
Availability of cleaning and disinfection products varies by country, and Mcube Technology is not
able to test the products in every market. Please contact your local service provider or Mcube
technology if you need more information.
Do not spray the disinfectant spray solution directly on the surface, recessed areas, crack, etc. of
the probe and the Charger.
CAUTION
This product may only be cleaned and disinfected by using the approved processes provided in this
manual. Cleaning and disinfection methods listed are recommended by Mcube Technology based on
compatibility with component materials.
Ensure that you follow the manufacturer's instructions for handling and disposing of the cleaning and
disinfection solutions provided in this manual.
BioCon-900S probe is intended to contact the abdominal skin to measure the bladder volume. For this
reason, cleaning and disinfecting the probe and Charger is an important procedure in using the system.
The following table shows the risk assessment for the probe including the Spauling's/CDC classification for
the minimum required disinfection level.
Component
Probe Head
Probe Except head
Charger
Compatibility
8.3.1
The following solutions have demonstrated cleaning or disinfection efficiency and material compatibility
with system components:
Metrex® CaviWipes™
Metrex® CaviCide™ Spray
Cleaning
8.3.2
Ensure the power cord is not plugged into wall outlet.
Using a dry paper towel or soft cloth, wipe any ultrasound gel, soils and dirt completely off the Charger or
Packaged
Use
Nonsterile
Reusable
Nonsterile
Reusable
Nonsterile
Reusable
TM
CUBEScan
BioCon-900S Operator's Manual
Spaulding's/CDC
classification
Noncritical
Noncritical
Noncritical
Disinfection level
Clean
Low
High
Page 44