CALIBRATION

8.9
BioCon-900S must be calibrated every 12 months to ensure accurate results. Calibration ensures accurate
alignment of BioCon-900S's internal system. You can take scan, even if calibration is not performed by the
designated date. But result can be compromised. There are following methods for the calibration.
CAUTION
Calibrate BioCon-900S when the battery is charged over 60%.
8.9.1 PHANTOM CALIBRATION
1) Place the CubeScan phantom on a flat
surface and open the cover of the phantom.
3) Place the phantom cover on the top of the
CubeScan phantom. Check if the cover is in a
stable and flat position.
CUBEScan
Phantom Calibration
2) Pour about 5ml of water or more on the center
surface of the phantom
4) Put the probe head into the probe holder firmly.
5) Turn the probe on and Tap the Setup icon (
on the bottom right side of the Home screen.
6) Tap the Next icon (
and select Phantom Calibration icon (
7) Tap the Scan icon (
You can cease the phantom calibration by
tapping the Cease icon (
(
) to return to Setup screen, if you want.
8) The calibration progress bar appears, when the
calibration is started. The progress bar indicates
the progress of calibration. It takes about
3minutes to complete the calibration. When the
calibration is completed, 100% sign is displayed
on the screen.
9) Tap the Ok icon (
date is updated. Tap the Exit icon (
to the Setup screen.
TM
BioCon-900S Operator's Manual
) on the Setup screen
) to start the calibration.
). Tap the Exit icon
) and the latest calibration
)
)
) to return