Display A Scan Results - Mcube CUBEscan BioCon-900S Operator's Manual

Table of Contents

Advertisement

Figure 27 Positioning Bladder
5) Tap the Scan icon(
scan. A Pre-scan image such as 2D ultrasonic image appears on the probe screen.
Figure 29 Pre-scan Screen
6) To locate the bladder, keep the probe contacted the abdomen. And tilt the probe slowly back and forth,
side to side. The position where the bladder image is the largest and most centered is the optimal place
to start the normal scan process. Locate the bladder along the centerline(1) on the Pre-scan screen. It
helps detect the optimal position during Pre-scan.
NOTE: Depth Line(3) is not displayed for male and female patient type.
7) If you want to stop the Pre-scan and return to the Information screen, tap the Exit icon(
8) Press the Scan Button
It takes 3 seconds to complete the normal scan Hold the probe steady while scanning. Movement of the
probe during scanning reduces the accuracy of measurement.
9) When you hear the beeps, the scan is complete.

DISPLAY A SCAN RESULTS

5.6
After a scan has been completed, the Scan Result screen appears. The Scan Result screen includes the
following information:
Current Urine Volume
Maximum Urine Volume
Patient Type & Tag
Scan Date & Time
Bladder Planes
Aiming Information
) on the Information screen or press the probe Scan Button
No.
1
Center Line
2
Bladder point: It helps to
easily recognize the bladder.
NOTE:
guaranteed that the pointer
shows the bladder in all
cases
3
Depth Line: Max detection
depth to measure a padiatric
patient's bladder volume.
4
Exit Icon
again or tap the Pre-scan screen. The progress bar on the screen appears.
TM
CUBEScan
BioCon-900S Operator's Manual
Figure 28 Probe Direction
Description
It
is
not
100%
to start Pre-
Figure 30 Normal Screen
).
Page 26