5 HOW TO USE

CHECKUP BEFORE USE

5.1
Before using BioCon-900S, ensure you are familiar with the device and its parts. If you use it for the first
time, we strongly recommend you to try your first bladder volume measurement on a patient with
moderately full bladder using Pre-scan. Pre-scan helps you detect the bladder and moderately full bladder
can be located more easily.
WARNING
If you use BioCon-900S in the presence of flammable anesthetics, explosion hazard exists.
Do not use the BioCon-900S on following cases:
1) Fetal use or pregnant patients.
2) Patients with ascites.
3) Patients with open or damaged skin.
4) Wounds in the suprapubic region.
CAUTION
If the oil leaks from housing and fluid gets in contact with your eyes, skin or clothing, flush the affected
area immediately with clean water and seek medical attention or see a doctor.

TURN SYSTEM ON/OFF

5.2
There are two power modes to system on/off the BioCon-900S. Loading screen depends on the power
mode and it shows during start-up.
Figure 9 Loading Screen (1)
SLEEP MODE
5.2.1
To preserve battery power, the probe goes into sleep mode by shutting itself down automatically when not
in use. You can set the time to go into sleep mode in the setup (2min/3min/5min). See the section "7.4
SLEEP MODE SETTING".
However, if a battery needs to be recharged immediately, Dialog box (
probe goes to sleep mode by shutting itself down. We strongly recommend that you should recharge the
battery immediately. See the section "4.2 CHARGE THE BATTERY".
Turn System on
CUBEScan
Figure 10 Loading Screen (2)
TM
BioCon-900S Operator's Manual
Figure 11 Home Screen
) appears shortly and then
Page 20