NORMAL SCAN SCREEN

6.5
Normal scan - Get 12-plane ultrasound images and calculate the residual urine in the bladder (3D
scanning). It takes 3 sec to complete the Normal scan. As bladder volume is calculated, the progress bar
increases until the scan is completed.
Figure 51 Normal Scan Screen

PRE-SCAN SCREEN

6.6
Pre-scan- 2D ultrasound images are displayed continuously in live image basis. This helps an operator
locate bladder position and predict the range of residual urine. And also the centerline is provided during
Pre-scan. Locate the bladder along the centerline on the screen. It helps detect the optimal position during
Pre-scan.
The Pre-scan is executed only when 'Pre-scan' setup is 'ON (
Normal scan is executed.
Figure 52 Pre-scan Screen (Default)
TM
CUBEScan
BioCon-900S Operator's Manual
No.
Description
1
Centerline.
2
Bladder point
Max detection depth to measure a pediatric urine
3
volume.
Cease the Pre-scan and return to the Information
4
screen.
Tap the Pre-scan screen, the Pre-scan is
5
executed.
)'. When 'Pre-scan' is 'OFF (
)', a