PNA Family Applications and Options (Continued)

Noise figure measurement summary

Overview
Description
Required options
Includes low-noise receiver
Includes filters for LO-harmonic
rejection
Includes source tuner bypass switch
Use standard receivers for noise figure
measurements
2
Vector noise calibration available using
ECal as tuner
Vector noise calibration using built-in
tuner
Recommended ECal for vector noise
calibration
3
Included tuner accessories for m-f ECal
Recommended adapter for f-f ECal
tuner (N4691D/93D/94D-F0F)
Scalar noise calibration available
S-parameter, conversion gain/loss measurements
Max freq (GHz)
NF measurements, amplifiers
Max frequency, using low-noise
receivers (GHz)
Max frequency, using standard
receivers (GHz)
2
Calibration accessories for low-noise
receivers
Calibration accessories for standard
receivers
1. For N5247B, the low-noise receiver and tuner works up to 50 GHz only.
2. External preamplifier and filter(s) required if DUT excess noise (gain plus noise figure) is < 30 dB up to
20 GHz, < 40 dB up to 50 GHz, or < 45 dB up to 67 GHz.
3. Only N4690 Series ECal modules are supported for use as an impedance tuner.
Find us at www.keysight.com
S93029B/A without Option 029
Noise figure application using standard receivers
only, for all PNA/PNA-X models
Not recommended with PNA Options 200, 210, 400,
410
No
No
No
Yes
Yes
No
3.5 mm N4691D-M0F (N5221/22/41/42/49B)
2.4 mm N4693D-M0F (N5224/25/44/45B)
1.85 mm N4694D-M0F (N5227/47B)
3.5 mm N5242-20169 cable, 85052-60013 m-f
adapter (N5221/22/41/42/49B)
2.4 mm N5245-20140 cable, 85056-60007 m-f
adapter
(N5224/25/44/45B)
1.85 mm N5247-20142 cable, 85058-60115 m-f
adapter (N5227/47B)
3.5 mm 85052-60014 m-m adapter
(N5221/22/41/42/49B)
2.4 mm 85056-60005 m-m adapter
(N5224/25/44/45B)
1.85 mm 85058-60113 m-m adapter (N5227/47B)
Yes
8.5/13.5/26.5/43.5/50/67
N/A
8.5/13.5/26.5/43.5/50/67
N/A
Cal kit or ECal, power meter
S93029B/A with Option 029
Noise figure application using standard or low-noise
receivers, for all PNA-X models
For N5241/42/49B, one of Options 21x, 22x, 41x or 42x
For N5244/45/47B, one of Options 22x or 42x
Yes
1
Yes
1
Yes
Yes
Yes
Yes, for N5244/45/47B
1
3.5 mm N4691D-M0F (N5241/42/49B)
Built-in for N5244/45/47B
3.5 mm N5242-20169 cable, 85052-60013 m-f adapter
(N5241/42/49B)
N/A for N5244/45B, due to built-in tuner
1.85 mm N5247-20142 cable, 85058-60115 m-f
adapter (N5247B, for use with external 67 GHz ECal)
3.5 mm 85052-60014 m-m adapter (N5241/42/49B)
N/A for N5244/45/47B, due to built-in tuner
Yes
8.5/13.5/26.5/43.5/50/67
8.5/13.5/26.5/43.5/50
8.5/13.5/26.5/43.5/50
Cal kit or ECal, 346C or 346C-K01 noise source or
power meter
Cal kit or ECal, power meter