Input Jack Sensing Test; Ir Port Verification - Fluke 1652C Calibration Manual

Electrical installation tester
Hide thumbs Also See for 1652C:
Table of Contents

Advertisement

Input Jack Sensing Test

This test determines if the input jack detect-circuits are operating correctly:
1. Install test leads into the L and PE UUT input jacks.
2. Set the rotary switch for Loop Z
3. Remove the test lead from the PE jack and note that LEAD appears on the display.
4. Reinsert the test lead into the PE jack and note that LEAD extinguishes.
5. Repeat steps 3 and 4 for the L jack.

IR Port Verification

You can do this verification procedure with Windows 98 or later. If you are
successfully using MET/CAL for reading data over the UUT IR port, it is
unnecessary to do this verification test.
To verify the UUT IR port:
1. Connect the Infrared Serial Cable to a COM port on the PC and the UUT IR port
adapter.
2. Press
.
3. Set the rotary switch to any function.
4. On the PC, go to Start, Programs, Accessories, and open HyperTerminal.
5. Setup the following:
Name: 165XB IR Port Test
Connect using: Com 1
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None
6. Enter the identification command <ID> followed by <Enter>.
Figure 3. 1654B LCD Tests
and use
to select L-PE.
I
Note
Electrical Installation Tester
Performance Tests
gad010.eps
19