8. Control (system test)
In menu „Control" (access with <Info> key) informations are displayed on 2 pages
(use cursor keys to scroll):
-software version
-serial number
-error amount
-operation hours since last service
-total operation hours
-operation hours pump
-distance of the piston
-date of the next recommended service
-phone number of the next service center
-accu voltage (charging status);
is displayed as a symbol in all menus:
Full charge
System test (check of instrument tightness):
1. introduce system test with <Enter>
2. lock connection „P high 1 +" with plug
3. start test with <Enter> (pressure is built up)
4. wait for result (approx. 2 min)
5. remove plug and leave with <Enter>
If the system is leaky,please contact your ecom partner.
Page 36
Half charge
Empty
Program version: V1.0 25.01.2012
Serial number : 4711
Error counter :
12
Operation hrs :
At a total
:
Pump
:
Cylinder (way) :
Next sevice
: 25.01.2013
Service tel.
: 02371/945-5
Further pages:
Programmversion: V1.0 25.01.2012
Geräte Nr.
: 4711
Battery
6.73 Volt
Systemtest:
Next sevice
: 25.01.2013
Service tel.
: 02371/945-5
Further pages:
ecom-DP3
8.45 std
18.75 std
0.17 std
13 m
Need help?
Do you have a question about the DP3 and is the answer not in the manual?
Questions and answers