Fig 3-37 : Ram's Hook Attachment

3.10.5 Calculating Tare Volume

The tare volume when using the adapter can be calculated by adding the fixed component
volume of 0.83 mL to the volume of the adjustable length tube. To calculate the volume of the
adapter tube, multiply the length of the tube by 0.054.