Flushing The 9705; Syringe Maintenance; Sensor Port Cleaning - Hach HIAC 9705 Operator's Manual

Liquid particle counter
Table of Contents

Advertisement

62 of 82

Flushing the 9705

5.3
1) Insert a bottle of deionized (DI) water into the Accu-Swirl or connect the Large Volume
2) From the Main Screen, press Flush. Flush as many as times as required by applicable
3) Remove water when finished.

Syringe Maintenance

5.4
The Cavro™ syringes provided for the syringe drive perform optimally when these guidelines
are followed:
• Do not run syringes for more than a few cycles without liquid in them.
• Thoroughly flush the syringe with distilled or deionized water after each use.
• If the plunger is removed from the syringe, wipe the plunger with alcohol before
• Clean the syringes periodically with a mild detergent. Allow the detergent to sit in the
• If the plunger does not move easily in the barrel, remove the plunger and wet the

Sensor Port Cleaning

5.5
The sensor port should be cleaned whenever large particles clog the cell.
1) Remove the clean-out port cap.
2) Retrieve a clean cell cleaning brush from the shipping container.
3) Slide the brush into the cleaning port, shown in
CAUTION:
Excessive pressure will result in brush and possible cell damage. If the brush will not go into
the cell, pull the brush out and inspect the cleaning tip for damage.
Operator Manual
Adapter (LVA) to a container of DI water.
SOPs. Flushing with at least 10 cm
replacement.
syringe for 30 minutes or more, then flush the syringe with distilled or deionized water for
at least 10 cycles.
plunger seal with deionized water.
encountered.
3
of clean fluid is recommended.
Figure 5-1
Maintenance Procedures - 9705
until resistance is
HIAC